PDA Technical Reports are highly valued membership benefits because they offer expert guidance and opinions on important scientific and regulatory topics and are used as essential references by industry and regulatory authorities around the world. These reports include terms which explain the material and enhance the reader’s understanding.

The database presented here includes the glossary terms from all current technical reports. The database is searchable by keyword, topic, or by technical report. Each definition provided includes a link to the source technical report within the  PDA Technical Report Portal.

(Please select "All" to restart a filtered Search)

  • All
  • 0
  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • A
  • B
  • C
  • D
  • E
  • F
  • G
  • H
  • I
  • J
  • K
  • L
  • M
  • N
  • O
  • P
  • Q
  • R
  • S
  • T
  • U
  • V
  • W
  • X
  • Y
  • Z
Air Removal Test
A test used to evaluate air removal and steam penetration in an empty sterilizer that is used for porous/hard goods load sterilization (e.g., Bowie-Dick Test, DART, Lantor Cube, Browns’ Test). (TR01) (TR 48)
Source:
Chamber Leak Test
A test conducted to evaluate possible air infiltration to the chamber under vacuum. [Synonym: Vacuum Leak Test] (TR1) (TR48)
Source:
Cool-Down Phase
The phase of a sterilization cycle that occurs after completion of the exposure phase. Parameters of a cool-down phase are typically defined in order to meet applicable user requirements for load cooling and drying. (TR01) The phase of a sterilization cycle that occurs after completion of the exposure phase. [Synonym: post-conditioning phase, slow exhaust phase, drying phase, equalization phase] (TR48) The phase of an SIP cycle that occurs after completion of the exposure phase. Parameters (e.g., time, temperature, pressure) of a cool-down phase are typically defined in order to meet applicable user requirements for system cooling and drying. (TR61)
Source:
Dryness Fraction
An absolute measure of the actual latent heat of a sample of steam relative to the theoretical latent heat of saturated steam. (TR01) (TR48)
Source:
Equilibration Time
The period that elapses between the attainment of the minimum exposure temperature at the reference measurement point (typically the drain) and the attainment of the sterilization temperature at all points within the load. This period is an indication of the ability to properly remove air and heat the load items; consequently, it is typically only evaluated by placing heat penetration probes in porous/hard goods loads. (TR01) (TR48)
Source:
F-Value (Lethality Factor)
A measurement of sterilization effectiveness, the F-value is the calculated equivalent lethality (using a specified z-value), in terms of minutes at a reference temperature (Tref), delivered by a sterilization cycle. (TR1) (TR3) (TR30) (TR48) (TR61)
Source:
Gravity Displacement
A sterilization process based on the principle that cold air within the chamber is heavier than the steam entering and will sink to the bottom of the chamber. As steam enters the chamber, air is pushed out the bottom drain and exits, with the condensate, through a steam trap. (TR01) (TR48) (TR61)
Source:
Heat-up Phase
The phase of a sterilization cycle that occurs prior to the exposure phase. Process parameters are developed for this phase in order to meet applicable user requirements for load conditioning (e.g., air removal and preheating.) (TR01) (TR3) (TR48) (TR61)
Source:
Liquid Load
A load consisting of closed containers of aqueous liquids. The sterilization of the container contents is achieved through transfer of energy through the container into the aqueous liquid. (TR01) (TR30) (TR48)
Source:
Load Zone
Area within the sterilization chamber where materials to be sterilized may be placed. (TR01) (TR3) (TR48)
Source:
Maximum Load
The maximum quantity or mass of items permitted in a sterilizer load. (TR01) The maximum quantity or mass of items permitted in a depyrogenation or sterilization load. (TR3) The maximum quantity or mass of products permitted in a validated sterilizer load. (TR30)The maximum quantity or mass of items permitted in a sterilizer load. (TR48)
Source:
Operating Parameters (Critical Parameters)
Values that are controlled and/or measured and are linked to safety and efficacy of a product or the process. Failure to meet a critical parameter should result in rejection of the load. (TR01) (TR3) (TR48) (TR51)
Source:
Operating Parameters (Key Parameters)
Values that are controlled and/or measured and are used to assure the ongoing “state of control” and consistency of runs. Failure to meet a key process parameter should result in an investigation with a documented rationale for the disposition of the load. (TR01) (TR3) (TR51) (TR48) Values that are controlled and/or measured and are used to assure the ongoing “state of control” of steam in place cycles. Failure to meet a key process parameter should result in an investigation. (TR61)
Source:
Saturated Steam
Steam that is at a temperature and pressure that corresponds to the vaporization curve of water. It is in a state of equilibrium between being a liquid and a gas, with no entrained liquid water. [Synonym: Dry Saturated Steam] (TR01) (TR48) (TR61)
Source:
Steam: Pure Steam
Collected condensate that complies with the Compendial requirements for Water for Injecton (WFI). [Synonyms: clean steam, high quality steam] (TR01) (TR61) (TR48)
Source:
Sterilization Cycle
A sequence of defined operating parameters (e.g., time, temperature and pressure) and conditions required to render an item sterile. (TR01) (TR30) (TR48) A sequence of defined operating parameters (e.g., time and temperature) required to render an item sterile. (TR3)
Source:
Temperature Distribution
Temperature measurement of the heating medium (e.g., forced hot air) across the chamber load zone. (TR01) (TR03) (TR30) (TR48)
Source: