PDA Technical Reports are highly valued membership benefits because they offer expert guidance and opinions on important scientific and regulatory topics and are used as essential references by industry and regulatory authorities around the world. These reports include terms which explain the material and enhance the reader’s understanding.
The database presented here includes the glossary terms from all current technical reports. The database is searchable by keyword, topic, or by technical report. Each definition provided includes a link to the source technical report within the PDA Technical Report Portal.
(Please select "All" to restart a filtered Search)
Refine Results
Filter By Technical Report Number
Filter By Technical Report Keyword
Filter By Technical Report Category
Air Removal Test
A test used to evaluate air removal and steam penetration in an empty sterilizer that is used for porous/hard goods load sterilization (e.g., Bowie-Dick Test, DART, Lantor Cube, Browns’ Test). (TR01) (TR 48)
Source:
Biological Qualification
A component of performance qualification that demonstrates, by use of biological indicators, that the required lethality (FBIO) is achieved consistently throughout the load. (TR1) (TR3) (TR30) A component of performance qualification that demonstrates, by use of biological indicators, that the required lethality (FBIO) or spore log reduction (SLR) is achieved consistently throughout the sterilized or sanitized portion of the SIP system. (TR61)
Source:
Chamber Cold Spot
The location(s) within the load zone that achieves the lowest process lethality (F0) and/or the lowest distribution temperatures during the sterilization process. (TR01)
Source:
Chamber Heat-Up Time
The elapsed time measured from the introduction of steam in the heat-up phase (“steam on”) to the point when the temperature of the heating medium within the chamber reaches the exposure temperature set point. (TR01)
Source:
Chamber Leak Test
A test conducted to evaluate possible air infiltration to the chamber under vacuum. [Synonym: Vacuum Leak Test] (TR1) (TR48)
Source:
Chemical Indicator
Test system that reveals change in one or more predefined process variables based on a chemical or physical change resulting from exposure to a process. (TR01) (TR30)
Source:
Chemical Integrator
A device that is designed to react in a quantitative manner to multiple sterilization variables, (typically, time and temperature and, in some instances, moisture). (TR01) (TR30)
Source:
F-Value (Lethality Factor)
A measurement of sterilization effectiveness, the F-value is the calculated equivalent lethality (using a specified z-value), in terms of minutes at a reference temperature (Tref), delivered by a sterilization cycle. (TR1) (TR3) (TR30) (TR48) (TR61)
Source:
F-Value (Lethality Factor) -- FBiological
A term used to describe the delivered lethality, measured in terms of actual kill of microorganisms on or in a BI challenge system. The FBiological-value is calculated as DT × LR, where DT is the D-value of the BI system at the reference temperature (T) and LR is the actual logarithmic reduction (log N0 – log NF) of the BI population achieved during the cycle. (TR1)
Source:
F-Value (Lethality Factor) -- FO
A term used when the specific reference conditions of Tref = 121.1°C and z = 10°C are used to calculate the equivalent lethality. For example, when the z-value of the BI is 10°C a cycle with an F(T=121.1°C, z=10°C), or F0, equal to 8 minutes is equivalent (in terms of delivered lethality) to a square wave cycle of 8 minutes at 121.1°C. A square wave cycle that provided an exposure of 25.9 minutes at 160deg;C would also yield an F0 of 8 minutes. (TR1)
Source:
F-Value (Lethality Factor) -- Fphysical
A term used to describe the delivered lethality calculated based on the physical parameters of the cycle. The FPhysical-value is the integration of the lethal rate (L) over time. The lethal rate is calculated for a reference temperature (Tref-) and z-value using the equation: L = 10(T-Tref- )/z. (TR1)
Source:
Liquid Load
A load consisting of closed containers of aqueous liquids. The sterilization of the container contents is achieved through transfer of energy through the container into the aqueous liquid. (TR01) (TR30) (TR48)
Source:
Load Zone
Area within the sterilization chamber where materials to be sterilized may be placed. (TR01) (TR3) (TR48)
Source:
Maximum Load
The maximum quantity or mass of items permitted in a sterilizer load. (TR01) The maximum quantity or mass of items permitted in a depyrogenation or sterilization load. (TR3) The maximum quantity or mass of products permitted in a validated sterilizer load. (TR30)The maximum quantity or mass of items permitted in a sterilizer load. (TR48)
Source:
Minimum Acceptable Cycle (MAC)
The minimum cycle conditions (in terms of delivered minimum lethality or minimum time and temperature) that would be considered acceptable. (TR01) (TR61)
Source:
Mixed Load
A load that contains multiple item item types representing various sterilization challenges. For example, some load items may have air removal challenges, while others pose a challenge due to their mass. (TR01)
Source:
Moist Heat
Steam, steam-air mixtures, and superheated water used for sterilization. (TR01)
Source:
Operating Parameters (Critical Parameters)
Values that are controlled and/or measured and are linked to safety and efficacy of a product or the process. Failure to meet a critical parameter should result in rejection of the load. (TR01) (TR3) (TR48) (TR51)
Source:
Operating Parameters (Key Parameters)
Values that are controlled and/or measured and are used to assure the ongoing “state of control” and consistency of runs. Failure to meet a key process parameter should result in an investigation with a documented rationale for the disposition of the load. (TR01) (TR3) (TR51) (TR48) Values that are controlled and/or measured and are used to assure the ongoing “state of control” of steam in place cycles. Failure to meet a key process parameter should result in an investigation. (TR61)
Source:
Overkill Design Approach
A sterilization design approach where minimal information is required about the product bioburden. A worst-case bioburden assumption is used to determine the delivered lethality needed to achieve a PNSU of 10-6 on or in the items being sterilized. When using this approach, the qualification program must demonstrate that both the FBIO and FPHYS are greater than 12 minutes. The required lethality may vary regionally. (Note: For typical SIP systems, the FPHYS will need to be greater than the FBIO.) (TR01) (TR3) (TR30) (TR61)
Source:
Parametric Release
A sterility release system based upon effective control, monitoring, documentation, and batch records review of a validated sterilization process cycle in lieu of release procedures based upon end-product sterility testing. (TR01) (TR3) (TR13) A sterility release program based on effective control, monitoring and documentation of a validated sterile-product manufacturing process where sterility release is based on demonstrated achievement of critical operational parameters and performance attributes in lieu of end-product sterility testing. (TR30)
Source:
Penetration Probe
A probe placed in contact with the load item or inside a container of liquid to measure the temperature of the load item or liquid. (TR01) A thermocouple placed in contact with the load item to measure the temperature of the load item. (TR3)
Source:
Physical Qualification
A component of performance qualification that demonstrates that predetermined physical requirements, including temperature distribution and heat penetration, are achieved consistently throughout the load. (TR01)(TR03)
Source:
Probability of a Non-Sterile Unit (PNSU)
The number that expresses the probability of occurrence of a non-sterile unit after exposure to a sterilization process. Within the pharmaceutical industry, a design end point better than or equal to the probability of one non-sterile unit in a million units is expected, i.e., PNSU ≤ 10–6. [Synonym: Steriliy Assurance Level (SAL)] (TR01)
Source:
Process Performance Qualification
Documented verification that a system is capable of consistently performing or controlling the activities of the processes it is required to perform or control, according to written and preapproved specifications, while operating in its specified operating environment. (TR01)
Source:
Product-specific Design Approach
A sterilization design approach that is based on the characteristics of the bioburden (on or in the load) and the heat sensitivity of the product that delivers the lethality needed to achieve a PNSU of 10-6 on or in the items to be sterilized. (TR01) (TR3) (TR30)
Source:
Saturated Steam
Steam that is at a temperature and pressure that corresponds to the vaporization curve of water. It is in a state of equilibrium between being a liquid and a gas, with no entrained liquid water. [Synonym: Dry Saturated Steam] (TR01) (TR48) (TR61)
Source:
Saturated Steam Process
A sterilization process, typically used for porous/hard goods loads, where the sterilizing medium is saturated steam. (TR01)
Source:
Steam Air Mixture (SAM) Process
A sterilization process in which the heating medium used to heat the load is in a mixture of air and steam that is typically used for liquid loads. This addition of air results in an air overpressure condition. (TR01)
Source:
Steam: Pure Steam
Collected condensate that complies with the Compendial requirements for Water for Injecton (WFI). [Synonyms: clean steam, high quality steam] (TR01) (TR61) (TR48)
Source:
Sterilization Cycle
A sequence of defined operating parameters (e.g., time, temperature and pressure) and conditions required to render an item sterile. (TR01) (TR30) (TR48) A sequence of defined operating parameters (e.g., time and temperature) required to render an item sterile. (TR3)
Source:
Sterilization Run
Execution of a sterilization cycle. (TR01)
Source:
Survivor Curve
Graphical representation of the inactivation of a population of microorganisms with increasing exposure to a microbicidal agent under stated conditions. (TR01) (TR3) (TR51) (TR61)