PDA Technical Glossary
PDA Technical Reports are highly valued membership benefits because they offer expert guidance and opinions on important scientific and regulatory topics and are used as essential references by industry and regulatory authorities around the world. These reports include terms which explain the material and enhance the reader’s understanding.
The database presented here includes the glossary terms from all current technical reports. The database is searchable by keyword, topic, or by technical report. Each definition provided includes a link to the source technical report within the PDA Technical Report Portal.
(Please select "All" to restart a filtered Search)
Refine Results
Filter By Technical Report Number
- TR 1: Validation: Moist Heat (13)
- TR 48: Moist Heat Sterilizer Systems (7)
- TR 30: Parametric Release (7)
- TR 61: Steam in Place (5)
- TR 51: Biological Indicators (3)
- TR 54: QRM:Manufacturing Operations (2)
- TR 69: Bioburden/Biofilm Management (2)
- TR 70: Cleaning/Disinfection Programs (2)
- TR 54-5: Quality Risk Management for the Design, Qualification, and Operation of Manufacturing Systems (2)
- TR 60: Process Validation (1)
- TR 13: Environmental Monitoring (1)
- TR 14: Validation: Protein Purification Chromatography (1)
- TR 82: Low Endotoxin Recovery (1)
- TR 42: Validation: Protein Manufacturing (1)
Filter By Technical Report Keyword
Filter By Technical Report Category
Biological Qualification
A component of performance qualification that demonstrates, by use of biological indicators, that the required lethality (FBIO) is achieved consistently throughout the load. (TR1) (TR3) (TR30) A component of performance qualification that demonstrates, by use of biological indicators, that the required lethality (FBIO) or spore log reduction (SLR) is achieved consistently throughout the sterilized or sanitized portion of the SIP system. (TR61)
Source:
Commissioning
A well planned, documented and managed engineering approach to the start-up and transfer of facilities, systems and equipment to the end-user that results in a safe and functional environment that meets established design and user requirement specifications. Commissioning precedes Qualification and includes three phases:
1. Inspection, testing, and regulation
2. Adjustment and setting of work
3. Functional testing (TR 3)
A prescribed number of activities designed to take equipment and systems from a static, substantially complete state to an operable state. (TR 48)
A well planned, documented, managed engineering approach to the start-up and turnover of facilities, systems, and equipment to the end-user, that results in a safe and functional environment that meets established design requirements and stakeholder expectations.(TR 54) (TR 54-5)
Source:
Depyrogenation
The destruction and/or removal of bacterial endotoxins. A depyrogenation process should demonstrate at least 99.9% or a 3-log endotoxin reduction. (TR3) Removal or destruction of pyrogens. (TR70)
Source:
F-Value (Lethality Factor)
A measurement of sterilization effectiveness, the F-value is the calculated equivalent lethality (using a specified z-value), in terms of minutes at a reference temperature (Tref), delivered by a sterilization cycle. (TR1) (TR3) (TR30) (TR48) (TR61)
Source:
F-Value (Lethality Factor)-- FH
A term used when the specific reference conditions of Tref = 160°C and z=20°C are used to calculate the equivalent lethality. For example, when the z-value of the BI is 20°C a process with an F(T=160°C, z=20°C), or FH, equal to 8 minutes is equivalent (in terms of delivered lethality) to a square wave process of 8 minutes at 160°C. A square wave process that provided an exposure of 45.2 minutes at 145°C would also yield an FH of 8 minutes. (TR3)
Source:
Leak Rate
Leak rate is the quantity of air leakage over time into the sterilizer chamber obtained while performing a chamber leak test. The leak rate should not exceed a level that will inhibit the sterilization process during air removal or vacuum drying stages. (TR03)
Source:
Lipopolysaccharide
A component of the cell wall of Gram-negative bacteria.(TR3)
Component of the outer cell wall of Gram-negative bacteria that is pyrogenic. (TR69)(TR82)
Source:
Load Zone
Area within the sterilization chamber where materials to be sterilized may be placed. (TR01) (TR3) (TR48)
Source:
Maximum Load
The maximum quantity or mass of items permitted in a sterilizer load. (TR01) The maximum quantity or mass of items permitted in a depyrogenation or sterilization load. (TR3) The maximum quantity or mass of products permitted in a validated sterilizer load. (TR30)The maximum quantity or mass of items permitted in a sterilizer load. (TR48)
Source:
Operating Parameters (Critical Parameters)
Values that are controlled and/or measured and are linked to safety and efficacy of a product or the process. Failure to meet a critical parameter should result in rejection of the load. (TR01) (TR3) (TR48) (TR51)
Source:
Operating Parameters (Key Parameters)
Values that are controlled and/or measured and are used to assure the ongoing “state of control” and consistency of runs. Failure to meet a key process parameter should result in an investigation with a documented rationale for the disposition of the load. (TR01) (TR3) (TR51) (TR48) Values that are controlled and/or measured and are used to assure the ongoing “state of control” of steam in place cycles. Failure to meet a key process parameter should result in an investigation. (TR61)
Source:
Overkill Design Approach
A sterilization design approach where minimal information is required about the product bioburden. A worst-case bioburden assumption is used to determine the delivered lethality needed to achieve a PNSU of 10-6 on or in the items being sterilized. When using this approach, the qualification program must demonstrate that both the FBIO and FPHYS are greater than 12 minutes. The required lethality may vary regionally. (Note: For typical SIP systems, the FPHYS will need to be greater than the FBIO.) (TR01) (TR3) (TR30) (TR61)
Source:
Parametric Release
A sterility release system based upon effective control, monitoring, documentation, and batch records review of a validated sterilization process cycle in lieu of release procedures based upon end-product sterility testing. (TR01) (TR3) (TR13) A sterility release program based on effective control, monitoring and documentation of a validated sterile-product manufacturing process where sterility release is based on demonstrated achievement of critical operational parameters and performance attributes in lieu of end-product sterility testing. (TR30)
Source:
Penetration Probe
A probe placed in contact with the load item or inside a container of liquid to measure the temperature of the load item or liquid. (TR01) A thermocouple placed in contact with the load item to measure the temperature of the load item. (TR3)
Source:
Physical Qualification
A component of performance qualification that demonstrates that predetermined physical requirements, including temperature distribution and heat penetration, are achieved consistently throughout the load. (TR01)(TR03)
Source:
Process Qualification
Documented verification that a system is capable of consistently performing or controlling the activities of the processes it is required to perform or control, according to written and preapproved specifications, while operating in its specified operating environment. (TR3)
Confirming that the manufacturing process as designed is capable of reproducible commercial manufacturing. (TR54) (TR60) (TR54-5)
Source:
Product-specific Design Approach
A sterilization design approach that is based on the characteristics of the bioburden (on or in the load) and the heat sensitivity of the product that delivers the lethality needed to achieve a PNSU of 10-6 on or in the items to be sterilized. (TR01) (TR3) (TR30)
Source:
Pyrogen
Any substance capable of eliciting a febrile (or fever) response upon injection or infection (as in endotoxin released in vivo by Gram-negative bacteria. (TR3) Fever-producing substance (TR69) A material that elicits a pyrogenic response (fever). (TR70)
Source:
Requalification
Periodic confirmation to demonstrate that equipment performance has not changed from its qualified state. (TR3)
Source:
Revalidation
Repeating partial or full validation of a process after a process change is implemented. Re-validation is change-based, not time-based. (TR14) (TR3) (TR42)
Source:
Routine Operational Process
Parameters that are specified for ongoing operations. The operational process is typically controlled to produce additional lethality over the qualified minimum parameters (i.e., time and temperature) in order to provide increased sterility assurance. (TR3)
Source:
Sterilization Cycle
A sequence of defined operating parameters (e.g., time, temperature and pressure) and conditions required to render an item sterile. (TR01) (TR30) (TR48) A sequence of defined operating parameters (e.g., time and temperature) required to render an item sterile. (TR3)
Source:
Sterilization Process
A process used to render a product free of viable organisms with a specified probability. (TR3)
Source:
Survivor Curve
Graphical representation of the inactivation of a population of microorganisms with increasing exposure to a microbicidal agent under stated conditions. (TR01) (TR3) (TR51) (TR61)
Source:
Water for Bacterial Endotoxin Test (BET)
Sterile Water for Injection or other water that shows no reaction with the specific bacterial endotoxin test reagent with which it is to be used, at the limit of sensitivity of such reagent. (TR3)