PDA Technical Glossary
PDA Technical Reports are highly valued membership benefits because they offer expert guidance and opinions on important scientific and regulatory topics and are used as essential references by industry and regulatory authorities around the world. These reports include terms which explain the material and enhance the reader’s understanding.
The database presented here includes the glossary terms from all current technical reports. The database is searchable by keyword, topic, or by technical report. Each definition provided includes a link to the source technical report within the PDA Technical Report Portal.
(Please select "All" to restart a filtered Search)
Refine Results
Filter By Technical Report Number
- TR 1: Validation: Moist Heat (7)
- TR 30: Parametric Release (5)
- TR 48: Moist Heat Sterilizer Systems (4)
- TR 61: Steam in Place (3)
- TR 51: Biological Indicators (2)
- TR 54: QRM:Manufacturing Operations (2)
- TR 70: Cleaning/Disinfection Programs (2)
- TR 13: Environmental Monitoring (2)
- TR 54-5: Quality Risk Management for the Design, Qualification, and Operation of Manufacturing Systems (2)
- TR 60: Process Validation (1)
- TR 69: Bioburden/Biofilm Management (1)
- TR 14: Validation: Protein Purification Chromatography (1)
- TR 42: Validation: Protein Manufacturing (1)
Filter By Technical Report Keyword
Filter By Technical Report Category
Biological Qualification
A component of performance qualification that demonstrates, by use of biological indicators, that the required lethality (FBIO) is achieved consistently throughout the load. (TR1) (TR3) (TR30) A component of performance qualification that demonstrates, by use of biological indicators, that the required lethality (FBIO) or spore log reduction (SLR) is achieved consistently throughout the sterilized or sanitized portion of the SIP system. (TR61)
Source:
Commissioning
A well planned, documented and managed engineering approach to the start-up and transfer of facilities, systems and equipment to the end-user that results in a safe and functional environment that meets established design and user requirement specifications. Commissioning precedes Qualification and includes three phases:
1. Inspection, testing, and regulation
2. Adjustment and setting of work
3. Functional testing (TR 3)
A prescribed number of activities designed to take equipment and systems from a static, substantially complete state to an operable state. (TR 48)
A well planned, documented, managed engineering approach to the start-up and turnover of facilities, systems, and equipment to the end-user, that results in a safe and functional environment that meets established design requirements and stakeholder expectations.(TR 54) (TR 54-5)
Source:
D-Value
The time in minutes required for a one-logarithm, or 90%, reduction of the population of microorganisms used as a biological indicator under specified lethal conditions. For dry-heat sterilization, the D-value should always be specified with a reference temperature, DT. For example, a biological indicator (BI) challenge system with a D 160°C=1.9 minutes, requires 1.9 minutes at 160°C to reduce the population by one logarithm. (TR3) The time in minutes at a specific temperature required to reduce the population of a specific microorganism by 90% [or one (1) log] in defined conditions [e.g., method of sterilization (dry heat versus steam), solute, or carrier]. (TR13)
Source:
Depyrogenation
The destruction and/or removal of bacterial endotoxins. A depyrogenation process should demonstrate at least 99.9% or a 3-log endotoxin reduction. (TR3) Removal or destruction of pyrogens. (TR70)
Source:
Dwell Time
The period that items are subjected to a given processing condition. [Synonym: Residence Time] (TR3)
Source:
Endotoxin Indicator (EI) for Depyrogenation
An article challenged with a vial of endotoxin (or a carrier spiked with endotoxin) designed for use in depyrogenation studies. The endotoxin (a purified lipopolysaccaride) is validated for use in or on an endotoxin indicator. The carrier is made from a material appropriate for the intended depyrogenation processes to which it will be subjected. The endotoxin on a carrier is added at a concentration sufficient to allow recovery of a minimum of 1000 USP endotoxin units/carrier. The endotoxin indicator would allow for accurate indication of at least a 3-log reduction in USP endotoxin units during depyrogenation process challenges. (TR3)
Source:
Operating Parameters (Critical Parameters)
Values that are controlled and/or measured and are linked to safety and efficacy of a product or the process. Failure to meet a critical parameter should result in rejection of the load. (TR01) (TR3) (TR48) (TR51)
Source:
Operating Parameters (Key Parameters)
Values that are controlled and/or measured and are used to assure the ongoing “state of control” and consistency of runs. Failure to meet a key process parameter should result in an investigation with a documented rationale for the disposition of the load. (TR01) (TR3) (TR51) (TR48) Values that are controlled and/or measured and are used to assure the ongoing “state of control” of steam in place cycles. Failure to meet a key process parameter should result in an investigation. (TR61)
Source:
Overkill Design Approach
A sterilization design approach where minimal information is required about the product bioburden. A worst-case bioburden assumption is used to determine the delivered lethality needed to achieve a PNSU of 10-6 on or in the items being sterilized. When using this approach, the qualification program must demonstrate that both the FBIO and FPHYS are greater than 12 minutes. The required lethality may vary regionally. (Note: For typical SIP systems, the FPHYS will need to be greater than the FBIO.) (TR01) (TR3) (TR30) (TR61)
Source:
Parametric Release
A sterility release system based upon effective control, monitoring, documentation, and batch records review of a validated sterilization process cycle in lieu of release procedures based upon end-product sterility testing. (TR01) (TR3) (TR13) A sterility release program based on effective control, monitoring and documentation of a validated sterile-product manufacturing process where sterility release is based on demonstrated achievement of critical operational parameters and performance attributes in lieu of end-product sterility testing. (TR30)
Source:
Process Qualification
Documented verification that a system is capable of consistently performing or controlling the activities of the processes it is required to perform or control, according to written and preapproved specifications, while operating in its specified operating environment. (TR3)
Confirming that the manufacturing process as designed is capable of reproducible commercial manufacturing. (TR54) (TR60) (TR54-5)
Source:
Product-specific Design Approach
A sterilization design approach that is based on the characteristics of the bioburden (on or in the load) and the heat sensitivity of the product that delivers the lethality needed to achieve a PNSU of 10-6 on or in the items to be sterilized. (TR01) (TR3) (TR30)
Source:
Pyrogen
Any substance capable of eliciting a febrile (or fever) response upon injection or infection (as in endotoxin released in vivo by Gram-negative bacteria. (TR3) Fever-producing substance (TR69) A material that elicits a pyrogenic response (fever). (TR70)
Source:
Requalification
Periodic confirmation to demonstrate that equipment performance has not changed from its qualified state. (TR3)
Source:
Revalidation
Repeating partial or full validation of a process after a process change is implemented. Re-validation is change-based, not time-based. (TR14) (TR3) (TR42)
Source:
Routine Operational Process
Parameters that are specified for ongoing operations. The operational process is typically controlled to produce additional lethality over the qualified minimum parameters (i.e., time and temperature) in order to provide increased sterility assurance. (TR3)
Source:
Sterilization Cycle
A sequence of defined operating parameters (e.g., time, temperature and pressure) and conditions required to render an item sterile. (TR01) (TR30) (TR48) A sequence of defined operating parameters (e.g., time and temperature) required to render an item sterile. (TR3)
Source:
Sterilization Process
A process used to render a product free of viable organisms with a specified probability. (TR3)
Source:
Thermometric Study
The utilization of independent temperature monitoring devices to determine a temperature profile within the load zone and analysis of the collected data. (TR3)
Source:
Water for Bacterial Endotoxin Test (BET)
Sterile Water for Injection or other water that shows no reaction with the specific bacterial endotoxin test reagent with which it is to be used, at the limit of sensitivity of such reagent. (TR3)